To create a template for running the custom pure dye plate: (continued)<br />

Step Action<br />

4 Apply pure dyes to the custom plate document as follows:<br />

a. Select the wells containing the custom dye.<br />

b. From the Setup tabbed page, click the Dyes drop-down list, and select the<br />

appropriate dye.<br />

The software applies the dye to the selected wells.<br />

c. Repeat steps 6 and 7 to configure the plate document with any additional<br />

custom dyes.<br />

Dyes drop-down list<br />

Custom dye added to selected<br />

wells of the plate document<br />

5 Save the custom pure dye plate document as a template file as follows:<br />

a. From the File menu, select Save.<br />

The Save dialog appears.<br />

b. Navigate to Program Files > Applied Biosystems > SDS 2.0 > Templates.<br />

The Templates directory appears within the Look in text field. By saving the<br />

template to the Templates folder, it becomes available from the Template<br />

drop-down list in the New Document dialog box.<br />

c. From the Files of type drop-down list, select <strong>ABI</strong> PRISM SDS Template Document.<br />

d. Click the File name text field, and type a name for the template document.<br />

e. Click Save.<br />

The software saves the plate document as a template file (*.sdt).<br />

6 Run the custom pure dye plate as explained in “Preparing and Running the<br />

Pure Dye Plate” on page 7-19.<br />

<strong>System</strong> Maintenance 7-23