To test the sensitivity of the plate sensor switch: (continued)<br />

Step Action<br />

9 Exit the Zymark Twister Software.<br />

a. Click Main Menu.<br />

The software displays the main menu.<br />

b. Click Quit Application.<br />

The software closes.<br />

10 Note A bug in the Zymark Twister Software can cause portions of the program to<br />

persist in memory even after the software has been closed. Because the Zymark<br />

Twister Software conflicts with the SDS software, the residual elements of the<br />

software must be closed within the Windows Task Manager before continuing.<br />

Confirm that the stack has closed by viewing the Task Manager.<br />

a. Press the Crtl + Alt + Del keys in unison.<br />

The Windows Security dialog box appears.<br />

b. Click Task Manager.<br />

The Task Manager dialog box appears.<br />

c. Confirm that the software has closed by looking for the Zymark Twister Software<br />

entry in the Task list. If the software is still running, click the software entry and<br />

click Close to exit the remaining software.<br />

d. From the File menu, select Exit.<br />

<strong>System</strong> Maintenance 7-31

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!