7-34 <strong>System</strong> Maintenance<br />

To align the input stack 1 (Zymark position 4): (continued)<br />

Step Action<br />

5 Using the Rotary Adjustment arrows, adjust the rotational position of the gripper so<br />

that it is centered over the input stack and will not contact the sides when lowered.<br />

♦ To move the plate handler arm clockwise, click the up arrow.<br />

♦ To move the plate handler arm counter-clockwise, click the down arrow.<br />

6 Using the Vertical Positioning commands, carefully lower the plate handler arm into<br />

the stack. Adjust the Rotary Adjustment value as needed to center the gripper<br />

within the stack.<br />

7 Once the gripper is centered within the stack, click Find Plate.<br />

The plate handler arm lowers upon the plate.<br />

Confirm the following:<br />

♦ the plate is in the middle of the gripper span<br />

♦ the plate sensor switch is contacting the plate<br />

♦ the gripper or plate do not contact the side of the stack<br />

8 Click Close Gripper.<br />

The gripper grips the plate between its fingers.<br />

9 Select Vertical Home.<br />

The plate handler raises the arm to its highest position. If the plate contacts the<br />

sides of the stack, re-adjust the rotary position of the plate handler arm until the<br />

plate moves freely in the stack.<br />

Note Contact between the plate and the stack or all stacks may be unavoidable.<br />

However, try to minimize the contact as much as possible.<br />

10 Using the Vertical Positioning commands, raise and lower the plate handler arm<br />

several times to check the alignment.<br />

11 Lower the plate handler arm to the bottom of the plate stack, click Rotary Offset,<br />

and click Yes.<br />

The software records the rotary position for the Zymark position 4 (input stack 1).<br />

12 Click Open Gripper.<br />

The gripper releases the plate.<br />

Up arrow (moves the arm clockwise)<br />

Down arrow (moves the arm counter-clockwise)