7-34 <strong>System</strong> Maintenance<br />
To align the input stack 1 (Zymark position 4): (continued)<br />
Step Action<br />
5 Using the Rotary Adjustment arrows, adjust the rotational position of the gripper so<br />
that it is centered over the input stack and will not contact the sides when lowered.<br />
♦ To move the plate handler arm clockwise, click the up arrow.<br />
♦ To move the plate handler arm counter-clockwise, click the down arrow.<br />
6 Using the Vertical Positioning commands, carefully lower the plate handler arm into<br />
the stack. Adjust the Rotary Adjustment value as needed to center the gripper<br />
within the stack.<br />
7 Once the gripper is centered within the stack, click Find Plate.<br />
The plate handler arm lowers upon the plate.<br />
Confirm the following:<br />
♦ the plate is in the middle of the gripper span<br />
♦ the plate sensor switch is contacting the plate<br />
♦ the gripper or plate do not contact the side of the stack<br />
8 Click Close Gripper.<br />
The gripper grips the plate between its fingers.<br />
9 Select Vertical Home.<br />
The plate handler raises the arm to its highest position. If the plate contacts the<br />
sides of the stack, re-adjust the rotary position of the plate handler arm until the<br />
plate moves freely in the stack.<br />
Note Contact between the plate and the stack or all stacks may be unavoidable.<br />
However, try to minimize the contact as much as possible.<br />
10 Using the Vertical Positioning commands, raise and lower the plate handler arm<br />
several times to check the alignment.<br />
11 Lower the plate handler arm to the bottom of the plate stack, click Rotary Offset,<br />
and click Yes.<br />
The software records the rotary position for the Zymark position 4 (input stack 1).<br />
12 Click Open Gripper.<br />
The gripper releases the plate.<br />
Up arrow (moves the arm clockwise)<br />
Down arrow (moves the arm counter-clockwise)