ABI Prism® 7900HT Sequence Detection System ... - OpenWetWare
ABI Prism® 7900HT Sequence Detection System ... - OpenWetWare
ABI Prism® 7900HT Sequence Detection System ... - OpenWetWare
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
Cleaning the Sample<br />
Block Wells<br />
7-12 <strong>System</strong> Maintenance<br />
To clean the sample block wells:<br />
Step Action<br />
1 Identify the contaminated wells as follows:<br />
a. If not already run, perform a background run as explained on page 7-13.<br />
b. Identify the contaminated wells on the sample block by following the procedure,<br />
“Isolating Sample Block Contamination” on page 8-8.<br />
2 Remove the sample block from the <strong>7900HT</strong> instrument as explained in “Removing<br />
the Sample Block” on page 7-5.<br />
3 Using the following figure as a guide, locate the suspected contaminated wells on<br />
the sample block.<br />
Warning<br />
GR2028<br />
Circuitry and connections to<br />
the instrument (Do Not Touch)<br />
Well A-1<br />
4 Pipet the appropriate volume of 10% bleach solution into each suspected<br />
contaminated well of the sample block module.<br />
♦ For a 96-well sample block, pipet 150 µL bleach solution to each well.<br />
♦ For a 384-well sample block, pipet 40 µL bleach solution to each well.<br />
5 Allow the sample block to sit for 3-5 min.<br />
6 Using a pipet, remove the bleach solution from the wells of the sample block.<br />
7 Rinse (pipet and remove) each contaminated well with 3 treatments of deionized<br />
water at the appropriate volume for the sample block.<br />
♦ For a 96-well sample block, rinse affected wells with 150 µL deionized water.<br />
♦ For a 384-well sample block, rinse affected wells with 40 µL deionized water.<br />
Note Absolute isopropanol can be substituted for water in the third treatment.<br />
8 Remove any remaining isopropanol or water from the wells of the sample block<br />
module.<br />
9 Replace the sample block as explained in “Replacing the Sample Block” on<br />
page 7-7.<br />
10 Run a background plate to confirm that the contamination has been removed.