Adding a Plate Document to the Plate Queue from the SDS Software<br />

Adding the Plate<br />

Document to the<br />

Plate Queue<br />

IMPORTANT A plate document must contain a bar code before it can be added to the plate<br />

queue. See page 4-19 for more information on configuring a plate document with bar code<br />

information.<br />

To add the plate document to the plate queue from the SDS software:<br />

Step Action<br />

1 From the SDS software, send the plate document to the queue as follows:<br />

a. Click the Instrument tab.<br />

The software displays the contents of the Instrument tabbed page.<br />

b. Click the Queue tab.<br />

The software displays the contents of the Queue tabbed page.<br />

2 Click Send to Queue.<br />

If… Then…<br />

the plate document was<br />

not saved previously<br />

the plate document was<br />

saved previously<br />

the Save As dialog box appears.<br />

a. From the Look in text field, navigate to a<br />

directory for the software to save the new file.<br />

b. From the Files of type drop-down list, select<br />

<strong>ABI</strong> PRISM SDS Single Plate (*.sds).<br />

c. Click the File name text field and either:<br />

– Type a name for the plate document file, or<br />

– Type or scan the bar code number of the plate<br />

into the text field.<br />

d. Click Save.<br />

The software saves the plate document.<br />

the software automatically saves the plate<br />

document.<br />

3 When prompted, click Yes to submit the document to the plate queue.<br />

Once a plate document has been added to the plate queue, the software locks the<br />

file preventing any changes from being made to it until the plate document has been<br />

run or removed from the queue.<br />

Note To release the plate document from the queue, launch the Automation<br />

Controller Software and remove the plate document from the queue as explained<br />

on page 4-35.<br />

4 Click OK to close the dialog confirming that the plate document has been added to<br />

the plate queue.<br />

5 From the File menu, select Close.<br />

The SDS software closes the plate document.<br />

6 Repeat the procedures in this chapter to create and add additional plates to the<br />

queue as needed.<br />

7 When finished creating plate documents, run the enqueued plates as explained in<br />

“Running Plates Using the Automation Controller Software” on page 4-34.<br />

Run Setup and Basic Operation 4-31