Setup Table Elements (continued)<br />

Number Contents Description<br />

Detector Definitions Element numbers 4 to 6 define the detectors that will appear in the well descriptions that<br />

follow in a later section. The detector definition consists of three sections: the declaration of<br />

the number of detectors, the detector list header, and the detector list.<br />

4 Number of<br />

Detectors<br />

Format:<br />

This line defines the total number of detectors on the plate.<br />

[ *** Number of Detectors number of detectors ]<br />

Example:<br />

*** Number of Detectors 5<br />

5 Detectors<br />

List Header<br />

Format:<br />

B-6 Importing and Exporting Plate Document Data<br />

This line contains the column headings for the Detector Definitions section of the setup table<br />

file that make the file easier to edit using a program such as Microsoft Excel.<br />

[ Detector Reporter Quencher Description Comments <br />

<strong>Sequence</strong> ]<br />

Example:<br />

Detector ReporterQuencher Description Comments <strong>Sequence</strong><br />

6 Detectors<br />

List<br />

Format for a single detector:<br />

The detector list consists of one or more lines displaying the information for each detector<br />

used on the plate document.<br />

♦ The Detectors List section must contain one line (or definition) for each detector present<br />

on the plate.<br />

♦ The number of lines in the Detectors List section must be equal to the number defined in<br />

the Number of Detectors section (see number 4 above).<br />

♦ Leave blank the Quencher Dye entry for detectors created for the SYBR Green I Dye or<br />

probes labeled with a non-fluorescent quencher.<br />

♦ The <strong>Sequence</strong> text value contains the name of the Allelic Discrimination marker<br />

associated with the detector. Because markers are used exclusively in allelic<br />

discrimination plate documents, assign <strong>Sequence</strong> text values to detector definitions of<br />

allelic discrimination setup table files only.<br />

[ detector name reporter dye quencher dye description <br />

comments sequence ]<br />

Example for an allelic discrimination setup table file:<br />

CYP 2C9*2.1 FAM PDAR CYP 2C9*2 Allele 1Example ProbeCYP 2C9*2<br />

CYP 2C9*2.2 VIC PDAR CYP 2C9*2 Allele 2Example ProbeCYP 2C9*2<br />

Example for an absolute quantification setup table file:<br />

GAPDH VIC GAPDH Probe Example Probe<br />

SYBR Green SYBR SYBR Green I Example Probe<br />

RNase P FAM TAMRA RNase P Probe Example Probe