Instrument<br />

Status Lights<br />

3-6 Getting Started<br />

The <strong>7900HT</strong> instrument contains three lights located on the lower-left side of the front<br />

panel to indicate the status of the instrument.<br />

Light/Appearance Status Action<br />

Green Solid The <strong>7900HT</strong> instrument is on<br />

and in idle state (ready to run)<br />

Flashing ♦ Interlocks are open and/or<br />

the scan head has not<br />

reached the safe position.<br />

♦ The instrument door is<br />

open.<br />

Orange Flashing The <strong>7900HT</strong> instrument is<br />

transmitting/receiving data<br />

to/from the computer<br />

(usually during a run).<br />

Solid If the light remains on during<br />

startup for more than 2 min:<br />

♦ The instrument did not boot<br />

properly, or<br />

♦ <strong>7900HT</strong> instrument has<br />

experienced a system<br />

failure<br />

Red Solid The <strong>7900HT</strong> instrument has<br />

detected a fatal problem.<br />

GR2010<br />

None<br />

This state indicates normal<br />

instrument function.<br />

None<br />

Red<br />

Orange<br />

Green<br />

Status lights<br />

Power button<br />

This state indicates normal<br />

instrument function.<br />

a. Check that the computer is<br />

turned on and connected to the<br />

instrument. (See page 2-11 for a<br />

diagram of instrument<br />

connections.)<br />

b. If so, turn off the instrument, wait<br />

for 30 sec, and then restart as<br />

explained on page 3-5.<br />

Turn off the instrument, wait for<br />

30 sec, and then restart.