Bit Error Rate TestSMIQ2.23.2 Block Error Rate Measurement (BLER)2.23.2.1 <strong>Operating</strong> MenuIn the operating menu, the BLER measurement is configured and the results are displayed.Fig. 2-1<strong>Operating</strong> menu for BLER measurementIn the upper section of the menu (display area), the measurement results are displayed, the way inwhich they have been obtained, and additional information on the applied signal. In the lower section,settings can be made.BLOCK ERRORRATE (BLER)Cyclic display of measured block error rate, the ratio of detected errors totransmitted data blocks.RECEIVED BLOCKSIEC/IEEE bus :BLER:R<strong>ES</strong>? Return value 3 ( of 7)Number of transmitted data blocks.ERRORSIEC/IEEE bus :BLER:R<strong>ES</strong>? Return value 1 ( of 7)Number of detected errors.TERMINATED BYIEC/IEEE bus :BLER:R<strong>ES</strong>? Return value 2 ( of 7)The displayed measurement result was obtained as follows:NUMBER OF RECEIVED BLOCKSThe defined number of data blocks was reached.NUMBER OF ERRORS The defined number of errors was detected.USERThe measurement was manually aborted by theuser.---- Measurement in progress.IEC/IEEE bus :BLER:R<strong>ES</strong>? Return value 4 ( of 7)1125.5555.03 2.378 E-9