SMIQFitting the OptionsFitting the option½ Plug the module into the appropriate slot for the FSIM2, lock it andfasten all screws.½ Open the air inlets at the housing frame by breaking out the safetyglass plate which belongs to the option.½ The following coaxial connections have to be made:FSIM2Module number1085.XXXXFSIM2Module number1114.XXXXCable from from toFSIM1W362 X361 X362 X362W364 X363 X364 X364W365 X365 X365 X365W366 X367 X368 X366W369 X370 X371 X369Note:Please store the remaining cables. They will be required iffurther options will be installed at a later stage or options willbe removed.½ The included adhesive label ’Option included’ is to be fixed at therear panel of the SMIQ.Calibrating theFading Simulator and theI/Q Modulators½ Warm-up the instrument for 2 hours.½ Call up menu UTILITI<strong>ES</strong> - CALIB - VECTOR MOD.Retrofit for 2-channel fadingCables W388 and W391 have to be repositioned for 2-channel fading.½ Unplug W388 from X368 of FSIM1 and plug to X368 (1085.XXXX)or X367 (1114.XXXX) of FSIM2.½ Unplug W391 from X371 of FSIM1 and plug to X371 (1085.XXXX)or X370 (1114.XXXX) of FSIM2.1125.5555.03 1.11E-7