SMIQUtilitiesCalibration ALC TABLEFor vector modulation or digital modulation, SMIQ is operated with the internal level control switched off.In mode ALC MODE - SAMPLE & HOLD, the level is recalibrated for each level or frequency setting. Inmode TABLE, the latter does not apply since the necessary input values are taken from a table. Thistable can be regenerated using LEARN TABLE . This is required to attain an optimum level accuracyfor the given operating temperature.Menu selection:UTILITI<strong>ES</strong> - CALIB - ALC TABLEFREQ100. 000 000 0 MHz LEVEL - 30.0 dBmFREQUENCYLEVELANALOG MODVECTOR MODDIGITAL MODDIGITAL STDLF OUTPUTSWEEPLISTMEM SEQUTILITI<strong>ES</strong>SYSTEMREF OSCPHASEPROTECTCALIBDIAGT<strong>ES</strong>TMOD KEYAUX I/OBEEPERINSTALLALLVCO SUMVECTOR MODLEV PR<strong>ES</strong>ETOSC REFLEVELALC TABLELEV ATTLEARN TABLEFig. 2-265Menu UTILITI<strong>ES</strong> - CALIB - ALC TABLELEARN TABLETriggers the generation of the new correction value for the function LEVEL-ALC-ALC OFF MODE TABLE.IEC/IEEE-bus command :SOUR:POW:ALC:TABL?1125.5555.03 2.421 E-9