SweepSMIQ2.26.7 LEVEL SweepMenu SWEEP - LEVEL offers access to settings for LEVEL sweep.Menu selection:SWEEP - LEVELFig. 2-247Menu SWEEP - LEVELSTART LEVELSTOP LEVELInput value of the starting level.IEC/IEEE-bus command :POW:STAR -30dBmInput value of the stop level.IEC/IEEE-bus command:SOUR:POW:STOP -10dBmCURRENT LEVELSTEPDWELLMODEEXECUTE SINGLE SWEEPIndication of the current level.<strong>Operating</strong> mode STEP: Input value of the level.Input value of the step width.IEC/IEEE-bus command :SOUR:SWE:POW:STEP 1dBInput value of the dwell time per stepIEC/IEEE-bus command :SOUR:SWE:POW:DWEL 15msSelection of the sweep operating mode.IEC/IEEE-bus commands :SOUR:POW:MODE SWE;:SOUR:SWE:POW:MODE AUTO;:TRIG:SOUR SINGStarts a single sweep run. This action to be executed is onlyindicated and is only effective if MODE SINGLE is selected.IEC/IEEE-bus command :TRIG1125.5555.03 2.398 E-9