GC and GC/MS<br />

228<br />

Columns for GC/MS<br />

Tips & Tools<br />

Agilent MS Certified Liners<br />

are lot-tested with MSD<br />

and FID for superior acid/base<br />

deactivation, response linearity<br />

and peak symmetry.<br />

Turn to page 153.<br />

Columns for GC/MS<br />

There is a rapidly increasing population of benchtop GC/MS instruments in analytical<br />

laboratories that analyze a broadening range of trace level, higher temperature samples.<br />

These samples require increasingly inert, lower bleed, higher temperature columns. In<br />

response to this growing need, Agilent Technologies deliberately designed several "ms"<br />

columns to chromatograph a broader range of low level samples and generate lower bleed<br />

even at higher temperatures.<br />

What makes an Agilent J&W low bleed column unique?<br />

Unique polymer chemistry and proprietary surface deactivation, both of which have<br />

contributed to columns that adhere to the tightest quality control specifications in the industry<br />

for bleed, inertness, selectivity and efficiency. Agilent J&W "ms" columns utilize special<br />

surface deactivation and siloxane chemistries which enhance the chromatographic<br />

performance of siloxane polymers.<br />

While some of the GC/MS phases utilize different polymer chemistries, their selectivity mimics<br />

the standard polysiloxane phases and offers the advantages of low column bleed and, in some<br />

cases, extended temperature ranges.<br />

The mass spectrum of septum bleed can look very much like GC column bleed, so the two are<br />

often confused. An easy way to tell the two apart: column bleed will be a rise in the baseline,<br />

not peaks. If you see bleed peaks, these generally come from lower quality septa or septa<br />

being used beyond their operating limits. To minimize septa contributions to background bleed<br />

use quality Agilent BTO, Long Life, or Advanced Green septa.<br />

www.agilent.com/chem/supplies