LC/MS Analysis of Angiotensin on Extend-C18<br />

Column: ZORBAX Extend-C18<br />

773700-902<br />

2.1 x 150 mm, 5 µm<br />

Mobile Phase: Acidic Conditions:<br />

A: 0.1% TFA in water<br />

B: 0.085% TFA in 80% acetonitrile (ACN)<br />

Basic Conditions:<br />

A: 10 mM NH4OH in water<br />

B: 10 mM NH4OH in 80% ACN<br />

Flow Rate: 0.2 mL/min<br />

Gradient: 15-50% B in 15 min<br />

Temperature: 35°C<br />

MS Conditions: Pos. Ion ESI- Vf 70V, Vcap 4.5 kV,<br />

N2- 35 psi, 12 L/min., 325°C<br />

Sample: 2.5 µL sample (50 pmol each)<br />

Angiotensin I, II, III<br />

A.<br />

Angiotensin I<br />

Max: 10889<br />

Low pH<br />

B.<br />

Angiotensin I<br />

Max: 367225<br />

TIC (200-1500 m/z)<br />

TIC (200-1500 m/z)<br />

5.0E6<br />

4.0E6<br />

3.0E6<br />

2.0E6<br />

1.0E6<br />

0<br />

5.0E7<br />

4.0E7<br />

3.0E7<br />

2.0E7<br />

1.0E7<br />

0<br />

0<br />

0<br />

2.5<br />

2.5<br />

AII + AIII<br />

5.0 7.5 10 12.5<br />

Time (min)<br />

AII<br />

5.499<br />

AI<br />

AIII<br />

8.477<br />

Reference: B.E. Boyes. Separation and Analysis of Peptides at High pH Using RP-HPLC/ESI-MS,<br />

4th WCBP, San Francisco, CA Jan 2000.<br />

www.agilent.com/chem/supplies<br />

9.621<br />

12.050<br />

13.261<br />

5.0 7.5 10 12.5<br />

Time (min)<br />

AI<br />

100<br />

80<br />

60<br />

40<br />

20<br />

0<br />

100<br />

80<br />

60<br />

40<br />

20<br />

0<br />

ZORBAX 300Å Extend-C18<br />

0<br />

+3<br />

227.1<br />

250.9<br />

340.3<br />

398.0<br />

433.0<br />

458.0<br />

Both small and large peptides demonstrate selectivity changes at high and low pH. At high pH, due to a change in charge, all three Angiotensins can be resolved. In<br />

addition, the spectral clarity of Angiotensin I is dramatically improved at high pH with the ammonium hydroxide mobile phase. The Extend-C18 column can be used for the<br />

analysis of small peptides at high pH as well.<br />

0<br />

433.0<br />

500<br />

+3<br />

500<br />

649.5<br />

+2<br />

659.7<br />

720.3<br />

767.3<br />

844.8<br />

649.0<br />

659.8<br />

+2<br />

1025.2<br />

1000<br />

1000<br />

1179.1<br />

1277.9<br />

1326.3<br />

1373.3<br />

1296.6<br />

1468.8<br />

m/z<br />

+1<br />

m/z<br />

LC30003<br />

LC and LC/MS<br />

695