Pumps<br />

www.agilent.com/chem/supplies<br />

Pump Supplies<br />

The Agilent 1100/1200 Series of HPLC modules has several different pump units to choose<br />

from: isocratic, quaternary, binary, capillary, nano, and preparative. The 1100/1200 was<br />

designed for ease of use – all pump maintenance parts can be accessed from the front by<br />

simply removing the front cover. Scheduled routine maintenance of these pumps based on<br />

your application is recommended to achieve maximum uptime.<br />

1100/1200 Pump Routine Maintenance Procedures<br />

• Replace the seals and pistons<br />

• Replace the PTFE frit<br />

• Replace the cartridge in the Active Inlet Valve<br />

• Clean the outlet ball valve<br />

• Clean or replace the solvent inlet frits<br />

Routine pump maintenance should be done on a regular basis to keep your Agilent<br />

1100/1200 system performing at its optimum. You may perform all the maintenance at once<br />

or perform maintenance on an as-needed basis. Some parts may need to be replaced more<br />

than others depending upon your application and solvent preparation procedures.<br />

Seal Wash<br />

The routine use of highly concentrated buffer solutions (100 mM) will reduce the life of the<br />

seals and pistons in your 1100/1200 Series pump. You can counteract the problem with one<br />

of our seal wash kits, which flushes the backside of the seal with a wash solvent;<br />

water/isopropanol (90/10) is recommended as the wash solvent. The continuous seal wash<br />

uses gravity to flow the wash solvent, and is compatible with 1100/1200 standard and<br />

capillary pumps. The active seal wash uses a peristaltic pump to flow the wash solvent, and is<br />

compatible with the 1200 Series pump and 1100 pumps with serial numbers DE40906378,<br />

DE40926032, DE40914884 and higher.<br />

LC and LC/MS<br />