Semi prep filter, 5064-8273<br />

High pressure semi prep filter, 5022-2165<br />

HPLC In-Line Filters<br />

www.agilent.com/chem/supplies<br />

Solvent Filtration<br />

Column inlet frit contamination can increase column back pressure and reduce efficiency.<br />

Microbore column blockages are a particular problem, due to the small diameter of the inlet<br />

frit. To prevent blockages, always use the appropriate filters in your LC system. Agilent offers<br />

two types of high pressure in-line filter kits for use with any HPLC system.<br />

Universal Solvent Filter<br />

Ideal for microbore, narrow-bore, high-speed<br />

or standard analytical columns.<br />

Universal solvent filters are installed between the LC pump and the injector, so particles from<br />

the solvent can be removed before they reach the injector. The filter assembly consists of a<br />

4.8 mm frit, two inserts and a two-piece holder. The frit is placed between the tapered edges<br />

of the inserts in such a way that the solvent is evenly distributed over the whole surface<br />

of the frit. This provides efficient filtration and extends the life of the frit.<br />

Low-volume Column Inlet Filter<br />

A high-capacity filter with built-in efficiency.<br />

This low-volume column inlet filter is positioned immediately before the LC column, so it can<br />

remove particles from both the injection system and the sample. With a frit diameter of only<br />

2.1 mm – plus tapered inserts – this filter minimizes external bandspreading while maximizing<br />

the filtration capacity.<br />

In Line Filters<br />

Low dispersion in-line filter Universal in-line filter<br />

Positioned immediately before<br />

the LC column<br />

Removes particles from the sample<br />

and injection system<br />

Minimizes external band spreading due<br />

to frit diameter of only 2.1 mm<br />

and tapered inserts<br />

Can be used with any microbore, high<br />

speed, or standard analytical columns<br />

Installed between the LC pump and the<br />

injector to remove particles from the solvent<br />

Uses a high capacity filter<br />

Frit is placed between the tapered edges<br />

of the inserts so the solvent is evenly<br />

distributed over the filtering frit<br />

LC and LC/MS<br />

495