Residual Solvents<br />

Column: DB-624<br />

123-1364<br />

60 m x 0.32 mm, 1.80 µm<br />

Carrier: Helium, 35-40 cm/sec, set to yield same<br />

RT for Hexane on all columns.<br />

Oven: 40°C Isothermal<br />

Sampler: Ambient Headspace<br />

Injection: Split ratio 1:8<br />

Detector: FID, 240°C<br />

Sample: 4 µL<br />

The three chromatograms to the right show how widely<br />

different the two other manufacturer’s columns are<br />

compared to the DB-624, the original USP G-43<br />

stationary phase column recommended for the analysis<br />

of these common organic volatile impurities in<br />

pharmaceutical samples.<br />

Top Chromatogram: DB-624<br />

Middle Chromatogram: Manufacturer A's "-624"<br />

Bottom Chromatogram: Manufacturer B's "-624"<br />

www.agilent.com/chem/supplies<br />

Life Science Applications<br />

2 4 6 8 10<br />

Time (min)<br />

Peak Numbers Critical Pair DB-624<br />

% Resolution*<br />

Manufacturer A's 624 Manufacturer B's 624<br />

2,3 ethanol and diethyl ether 100% 0% 0%<br />

5,6 ethyl acetate and 2-butanone 95% 38% 0%<br />

7,8 2-butanol and chloroform 60% 0% 60%<br />

9,10 benzene and isooctane 100% 0% 0%<br />

*Resolution calculated as follows: %R = 100% x valley height/average of peak height 1 + peak height 2<br />

Tips & Tools<br />

pA FID1 A1 100<br />

80<br />

60<br />

40<br />

20<br />

pA FID1 A1 100<br />

80<br />

60<br />

40<br />

20<br />

pA FID1 A1 100<br />

80<br />

60<br />

40<br />

20<br />

1<br />

1<br />

2 4 6 8 10<br />

1<br />

2<br />

2<br />

2<br />

3<br />

3<br />

3<br />

4<br />

2 4 6 8 10<br />

4<br />

4<br />

1. Methanol<br />

2. Ethanol<br />

3. Diethyl ether<br />

4. Hexane<br />

5. Ethyl Acetate<br />

6. 2-Butanone<br />

7. 2-Butanol<br />

8. Chloroform<br />

9. Benzene<br />

10. Isooctane<br />

5 6<br />

7 8<br />

5 6<br />

5, 6<br />

7 8<br />

For the latest residual solvent application for USP 467, request publication number 5989-8085EN.<br />

7, 8<br />

9<br />

9<br />

9 10<br />

10<br />

10<br />

GCLS029<br />