GC and GC/MS<br />

GC Column Troubleshooting and Maintenance<br />

Excessive Baseline Noise<br />

Possible Cause Solution Comments<br />

Injector contamination Clean the injector; replace liner, gold seal Try a condensation test; gas lines may also<br />

need cleaning<br />

Column contamination Bake-out the column Limit the bake-out to 1-2 hours<br />

Solvent rinse the column Only for bonded and cross-linked phases<br />

Check for inlet contamination<br />

Detector contamination Clean the detector Usually the noise increases over time and<br />

not suddenly<br />

Contaminated or low quality gases Use better grade gases; also check for expired<br />

gas traps or leaks<br />

Usually occurs after changing a gas cylinder<br />

Column inserted too far into the detector Reinstall the column Consult GC manual for proper insertion distance<br />

Incorrect detector gas flow rates Adjust the flow rates to the recommended values Consult GC manual for proper flow rates<br />

Leak when using an MS, ECD, or TCD Find and eliminate the leak Usually at the column fittings or injector<br />

Old detector filament, lamp or<br />

electron multiplier<br />

Replace appropriate part<br />

Septum degradation Replace septum For high temperature applications use an<br />

appropriate septum<br />

350 www.agilent.com/chem/supplies<br />

GC Column Troubleshooting and Maintenance<br />

These at-a-glance troubleshooting tables will help you pinpoint and fix the most common GC<br />

column problems.<br />

Please consult Agilent's J&W GC Column Installation Guide (part number 830-0120) and<br />

the Agilent J&W GC Column Selection Guide (publication number 5989-6159EN) for more indepth<br />

information, including:<br />

• Comprehensive column selection principles<br />

• Maintenance procedures, including column installation, conditioning, testing, and storage<br />

• Strategies for increasing your productivity<br />

• The most current method development procedures<br />

To request your copy of the Agilent J&W GC Column Selection Guide, visit<br />

www.agilent.com/chem/guides.<br />

The Agilent J&W GC Column Installation Guide comes standard with every Agilent J&W GC<br />

column. You can also request a copy by contacting your local Agilent Representative or<br />

Agilent Authorized Distributor.