For PDA systems with 57X08 <strong>and</strong> 57X18 type photomulitpliers, the PM<br />

high voltage will be switched of in case of excessive PM anode current. It is<br />

not possible to reduce the high voltage gradually, because this would<br />

require a new phase calibration for each new setting of the PM high voltage.<br />

The Scope display’s triggering <strong>and</strong> scaling are explained in Chapter 5 –<br />

Group poperties.<br />

The CPU load can be checked by the Windows Task Manager’s<br />

Performance tab. If it is near 100% during acquisition, it is recommended to<br />

set Scope display Off, <strong>and</strong> do disable as many objects as possible during<br />

acquisition. See chapter 5 for details.<br />

13. Before taking data, you should open the Phase plot graph to check the<br />

alignment of the receiving optics with the transmitting optics. Right-click on<br />

<strong>BSA</strong> application in the project explorer, click on New <strong>and</strong> select phase<br />

plot. During a measurement, the following graph appears:<br />

For further details about the Phase Plot, please refer to chapter 5.<br />

14. To display a velocity <strong>and</strong> a diameter distribution histograms, right-click on<br />

the <strong>BSA</strong> Application icon in the project explorer <strong>and</strong> select New. Select<br />

Histogram from the dialog. A new window appears in the workspace.<br />

4-60 <strong>BSA</strong> <strong>Flow</strong> <strong>Software</strong>: The user interface