6.7.2 Hardware connections<br />

If this option was not purchased at the same time as the processor, the<br />

processor must be returned to the distributor for installation of the hardware.<br />

The <strong>BSA</strong> processor software must be updated to version 2.30 or later. This is<br />

done by downloading from the CD comprising <strong>BSA</strong> <strong>Flow</strong> software <strong>and</strong><br />

processor software, or from Dantec Dynamics’ web site.<br />

Slot for 2 nd Analog/Digital input board<br />

Analog input connector with BNC<br />

adapter attached<br />

Digital input connector<br />

Figure 6-72 <strong>BSA</strong> analog/digital input: rear panel connections<br />

Analog inputs can be connected to the analog input connector shown in<br />

Figure 6-72 via the adapter shown in Figure 6-71.<br />

Analog inputs Up to 4 analog signals can be sampled on the input connectors labelled<br />

Channel 1 to 4.<br />

Signal Levels The Analog input option supports bipolar signals:<br />

Bipolar: ±10 V<br />

Synchronisation Synchronisation of A/D-data with the LDA-data is done internally in the<br />

processor.<br />

The analog input board must be associated with a coincidence group.<br />

If two boards are used, they can be associated with different coincidence<br />

groups.<br />

The analog data are acquired when bursts are detected in the coincidence<br />

group.<br />

For set-up please refer to the next section.<br />

6-86 <strong>BSA</strong> <strong>Flow</strong> <strong>Software</strong>: Options <strong>and</strong> Add-ons