Set-up of the remote PC<br />

5.8 Coincidence<br />

In the address field above, type only the first part of the name, without the<br />

domain name.<br />

To activate the traverse server, run the VtravServer.exe program located in<br />

the <strong>BSA</strong> <strong>Flow</strong> <strong>Software</strong> folder. This will show the traverse server icon in<br />

the lower left corner of the screen.<br />

Right-click on the icon – select Settings <strong>and</strong> check the Use TCP/IP Network<br />

box:<br />

Click OK.<br />

The traverse system is now ready for remote operation.<br />

Due to different sensitivity of the various receivers, a seeding particle<br />

passing through the outskirts of the measuring volume may sometimes<br />

generate a velocity sample on one channel, without simultaneously<br />

generating a sample on the other(s).<br />

Most calculations involving two or more velocity components require that<br />

the velocity samples in question are coincident (i.e. simultaneous). Consider<br />

coordinate transformation as an example: Calculating an orthogonal velocity<br />

sample with components u, v <strong>and</strong> w from “skew” components u1, u2 <strong>and</strong> u3<br />

obviously require that the latter are simultaneous. Otherwise the calculated<br />

(orthogonal) velocity sample would have no meaning, since it could not be<br />

related to any specific point in time. Similar considerations apply when<br />

calculating cross-moments (Reynolds stresses).<br />

Note the 62N <strong>BSA</strong> can be set up to provide coincident data by placing velocity<br />

channels in the same group. This is hereafter referred to as hardware<br />

coincidence.<br />

If velocity channels are placed in separate groups, software coincidence<br />

can be applied by adding a Coincidence object as described in this section.<br />

Coincidence filtering is a child of a data source object. Possible parent<br />

objects are <strong>BSA</strong> F/P Application <strong>and</strong> Import velocity files as shown in<br />

Figure 5-40.<br />

5-48 <strong>BSA</strong> <strong>Flow</strong> <strong>Software</strong>: Project explorer objects