8.3.2 Phase plot<br />

Figure 8-5: Calibration. PM balance.<br />

The displaid values in the warning window are the level ratio between each<br />

PM <strong>and</strong> the level reach by the strongest PM.<br />

In above case, the strongest Photomultiplier is PM2 <strong>and</strong> the weekest PM is<br />

PM 1. The ratio between Photomultiplier 2 <strong>and</strong> Photomultiplier 1 is higher<br />

than 2. The solution is either increasing the high voltage of channel 1 or<br />

reducing high voltage of channel 2.<br />

For the description of the Phase plot please refer to 5.14-Phase plot chapter.<br />

Note With low particle Data rate, the refreshment of the plot is not as fast as at<br />

high particles data rate.<br />

Figure 8-6: example of Phase Plot.<br />

<strong>BSA</strong> <strong>Flow</strong> <strong>Software</strong>: Trouble shooting 8-7