10. ptical LDA System properties include the wavelength <strong>and</strong> other optical<br />

parameters.<br />

The optical parameters that you defined in the Dantec Device<br />

Configuration will appear in the list. If necessary, you can modify them.<br />

The default value of Exp<strong>and</strong>er ratio is 1 corresponding to no beam<br />

exp<strong>and</strong>er.<br />

The nominal Beam diameter, Exp<strong>and</strong>er ratio <strong>and</strong> Beam spacing are<br />

specified in the user’s guide for the optics.<br />

The Frequency shift is 40000000 Hz for Dantec Dynamics Fiber<strong>Flow</strong> <strong>and</strong><br />

<strong>Flow</strong>Lite optical systems, <strong>and</strong> 80000000 Hz for <strong>Flow</strong>Explorer optical<br />

systems.<br />

Note The software computes the effective beam spacing at the front lens as Beam<br />

Spacing multiplied by Exp<strong>and</strong>er Ratio<br />

Alternatively, the beam spacing at the exit of the beam exp<strong>and</strong>er can be<br />

used. In that case, set the Exp<strong>and</strong>er Ratio to 1, <strong>and</strong> the Beam Spacing to<br />

the one at the exit of the beam exp<strong>and</strong>er. In the Beam diameter field, enter<br />

the beam diameter at the front lens, i.e. the beam diameter before beam<br />

exp<strong>and</strong>er(s) multiplied with the total beam expansion.<br />

To monitor the Doppler signals, right click on Processor, Group 1 or LDA 1<br />

in the <strong>BSA</strong> Application window, <strong>and</strong> select System Monitor from the list.<br />

An “always on top” window appears. An oscilloscope-like display will<br />

appear in this window. This shows a number of samples of the filtered <strong>and</strong><br />

digitised Doppler bursts.<br />

In addition to the oscilloscope display it includes a Group pane <strong>and</strong> a<br />

Channel pane showing the anode current of the photo-multiplier(s) <strong>and</strong><br />

synchronisation pulses if available.<br />

The panes can be collapsed individually by clicking on the button in their<br />

upper right corner.<br />

4-50 <strong>BSA</strong> <strong>Flow</strong> <strong>Software</strong>: The user interface