Methods to determine whether the LDA measuring volume is in position<br />

Note For all the below procedures, it is important to protect the photomultiplier<br />

tubes. Run the laser at low power <strong>and</strong> make sure that the PM current limit<br />

property of the processor has been correctly set.If the PM current saturates,<br />

reduce the PM high voltage.<br />

The main difficulty is normally to determine whether the measurement<br />

volume’s longest dimension is centered correctly. The below methods can be<br />

used for this. The pinhole/power meter method is good for determination of<br />

all three coordinates.<br />

Pinhole/power meter method<br />

A precise method is to mount a pinhole matched to the diameter of the<br />

measuring volume within the flow rig, <strong>and</strong> accurately measure its position<br />

relative to the laboratory origin. The LDA measuring volume must then be<br />

moved to the position where maximum light power from all beams pass<br />

through the pinhole. This method is highly repeatable, if a power meter is<br />

used to measure the power of the beams after the pinhole.<br />

Window surface/photomultiplier current method<br />

Another often used method is to use a window surface as the reference for<br />

the LDA measuring volume: if there is a window between the LDA optics<br />

<strong>and</strong> the flow, <strong>and</strong> the LDA measuring volume can be traversed to the<br />

window position, the detected Doppler signal from the wall or the PM anode<br />

current will reach a maximum, when the center of the measuring volume is<br />

at the surface of the window.<br />

Model surface/photomultiplier current method<br />

Similar to the window surface method, but if the surface is not transparent,<br />

stronger reflections occur, <strong>and</strong> it may be necessary to reduce the PM high<br />

voltage more. The automatic protection of the PM may further reduce the<br />

high voltage. In this case, reduce the laser power if possible until the PM<br />

current is below the allowed limit.<br />

5.7.7 Saving traverse properties for use in other projects<br />

To ensure that new projects use the traverse properties that you have defined,<br />

it is necessary to perform the steps below.<br />

The template function supports the properties of the Lightweight,<br />

Lwh<strong>and</strong>box <strong>and</strong> Generic traverse systems, but if you define new projects<br />

without aplying a template, the traverse properties will be the system default<br />

values.<br />

To save traverse settings in such a way that they are automatically used in<br />

future projects, you must right-click the traverse server icon <strong>and</strong> select<br />

Save... This opens a dialog in which you can rename the traverse server to<br />

any name that you wish, e.g. Lab Traverse.<br />

In the Dantec Device Configuration, select Custom Driver, <strong>and</strong> type in the<br />

name you gave the traverse:<br />

5-42 <strong>BSA</strong> <strong>Flow</strong> <strong>Software</strong>: Project explorer objects