6.2.7 Specifications<br />

TTL-level inputs<br />

BNC<br />

Differential synchronisation input<br />

1k00<br />

1n<br />

74HCT245<br />

Figure 6-39 TTL level synchronisation input<br />

Input TTL-level<br />

Nominal low-level: 0V ≤ VIL ≤ 0.8 V<br />

Nominal high-level: 2V ≤ VIH ≤ 5 V<br />

Input filtering RC network, 1 kΩ <strong>and</strong> 1nF (time constant<br />

1µsec.)<br />

Input buffer 74HCT245 type<br />

Pulse duration, high level ≥ 2 µsec<br />

Pulse duration, low level ≥ 2 µsec<br />

Input protection Diodes to ground <strong>and</strong> +5V; 1 kΩ series<br />

protection resistors<br />

Safe input range: ± 10 V continuous, instument power off<br />

Table 6-4 TTL-input specifications<br />

- 10 V to +15 V continuous, instument<br />

power on<br />

± 20 V transient<br />

higher levels may damage the unit<br />

Please observe that there are no pull-up or pull-down resistors on the inputs.<br />

This means that the logic level of an open input is undefined.<br />

An input, which has been selected as a source for one or more functions,<br />

must be driven to a valid logic level.<br />

A +5V output is available at S1 (25 way D-sub connecor). This voltage can<br />

be used for providing pull-up if for example an input must be driven by an<br />

open-collector type of output<br />

IN+<br />

IN-<br />

R50 100R<br />

R52 100R<br />

C114<br />

2n2<br />

6-38 <strong>BSA</strong> <strong>Flow</strong> <strong>Software</strong>: Options <strong>and</strong> Add-ons<br />

+<br />

-<br />

26C32