Avoid overloading the photomultipliers.<br />

For LDA systems with 57X08 <strong>and</strong> 57X18 type photomulitpliers, the<br />

processor’s built-in protection circuit will gradually reduce the PM high<br />

voltage in case of excessive PM anode current, but it is recommended to<br />

avoid overload situations.<br />

In case of measurements between the blades of an impeller the PM current<br />

will be high when an impeller blade passes through the measurement<br />

volume, <strong>and</strong> this will overload the PM current for a short period, which is<br />

acceptable. However, if possible, it is recommended to establish a laser<br />

shutter system which blocks the laser beams during passages of solid<br />

surfaces through the measurement volume<br />

The Scope display’s triggering <strong>and</strong> scaling are explained in Chapter 5 –<br />

Group poperties.<br />

The CPU load can be checked by the Windows Task Manager’s<br />

Performance tab. If it is near 100% during acquisition, it is recommended to<br />

set Scope display Off, <strong>and</strong> do disable as many objects as possible during<br />

acquisition. See chapter 5 for details.<br />

12. . Before taking data, a result display should be defined. To display a<br />

velocity distribution histogram, right-click on the <strong>BSA</strong> Application icon<br />

in the project explorer <strong>and</strong> select New. Select Histogram from the<br />

dialog. A new window appears in the workspace.<br />

The axes will automatically rescale when you start acquiring data.<br />

13. To start acquisition, click on <strong>BSA</strong> Application in the project explorer,<br />

<strong>and</strong> click on the Run button in the tool bar (or press the F5 button on<br />

the keyboard).<br />

The histogram will be updated during the measurement. With 2000<br />

samples selected, all data is usually sent in a single block, so the<br />

histogram updates once only. The update rate depends on the data rate.<br />

With high data rates <strong>and</strong> a large number of samples, you will see several<br />

updates during an acquisition.<br />

When the acquisition is completed, the “Run” button is clear again. If the<br />

System monitor is open, either the Samples or Acquisition bar graph is at<br />

full scale.<br />

Under the menu “Tools - Options - General” you can select “Feedback<br />

with sound”. With this option active, you will hear a sound from the PC<br />

when acquisition is completed.<br />

4-52 <strong>BSA</strong> <strong>Flow</strong> <strong>Software</strong>: The user interface