Beam spacing The spacing between the beams in the selected beam system, before the<br />

beam exp<strong>and</strong>er. This figure is multiplied by the exp<strong>and</strong>er ratio to find the<br />

resulting beam spacing at the exit of the beam exp<strong>and</strong>er or entry to the front<br />

lens. Used for calculation of the fringe spacing in <strong>and</strong> dimensions of the<br />

measurement volume.<br />

Frequency shift: The specified frequency is the Bragg cell frequency. This value can positive<br />

or negative according to frequency shift direction. The normal value is<br />

positive. In this case, a positive flow velocity will produce a photo-multiplier<br />

signal frequency higher than the shift frequency.<br />

Negative is used if the optical system is set up to reduce the signal frequency<br />

by an amount corresponding to the shift frequency. It is typically the case if<br />

the velocities are so high that they would produce out-of-range signal<br />

frequencies for the processor, if a positive frequency shift was chosen.<br />

5.5 Optical Calibrated LDA System properties<br />

Dantec Dynamics delivers calibration data on a CD-ROM for certain optical<br />

systems.<br />

The calibration data are automatically installed when inserting the CD-ROM,<br />

if Auto-run is enabled.<br />

The calibration data includes the calibration date, the calibrated fringe<br />

spacing, <strong>and</strong> nominal values for the optical parameters (wavelength, focal<br />

length, beam spacing, beam diameter). The relevant parameters are<br />

transferred to the Device Configuration, so that the software uses these when<br />

new projects are created.<br />

To see the parameters, start the device Configuration :<br />

Select Optical Calibrated LDA System:<br />

5-28 <strong>BSA</strong> <strong>Flow</strong> <strong>Software</strong>: Project explorer objects