Alignment<br />

of the eyepiece The focusing of the image onto the slit-shaped spatial filters can be<br />

inspected via an alignment eyepiece. The alignment eyepiece is a lens<br />

magnifying the image of the focusing plane on the fibre end of fiber V1 so<br />

that the DualPDA Probe can be accurately aligned with the beam<br />

intersection volume.<br />

Front lenses A range of front lenses is available. The front lens is a separate unit not<br />

included with the probe. All front lenses for the DualPDA receiving optics<br />

can be used on the 60X60, 60X61, 60X62, 60X63, 60X64, 60X65, 60X66 <strong>and</strong><br />

60X67 Fiber<strong>Flow</strong> Probes <strong>and</strong> vice versa.<br />

Size range The measurable size range is determined by the following parameters:<br />

• Beam intersection angle of the transmitting optics (defined by the<br />

beam separation <strong>and</strong> the front lens focal length).<br />

• Focal length of the receiving front lens.<br />

• Aperture plate.<br />

• Selected polarization.<br />

• Scattering angle.<br />

Figure 7-42: The image of the beam crossing on the spatial filter.<br />

The 57X80 DualPDA Probe is compatible with the 60X60 <strong>and</strong> 60X61 Fiber-<br />

<strong>Flow</strong> probes concerning the use of beam exp<strong>and</strong>ers, side-looking section,<br />

probe supports, etc.<br />

Aperture plates The aperture plate defines the receiving apertures. The size range can be<br />

selected by the choice of aperture plate, without changing the focal length of<br />

the front lens or the scattering angle.<br />

Three st<strong>and</strong>ard plates are included with the probe (Figure 7-43, Figure 7-44),<br />

A, B <strong>and</strong> C. The effective spatial (azimuthal) position of each segment, <strong>and</strong><br />

thus the particle size sensitivity <strong>and</strong> range, depends on which plate is used.<br />

The aperture plates are exchangeable.<br />

<strong>BSA</strong> <strong>Flow</strong> <strong>Software</strong>:Reference guide 7-47