alloy. In contrast, that of average 40.257με (4.0%) were observed at the part of low<br />

elastic modulus that circled. Namely, that compared with the femoral nail made of a<br />

titanium alloy, the original PEEK femoral nail has been confirmed to have occurred a<br />

large strain (figure 5). Figure 6 shows the strain distribution of bone model in the<br />

original PEEK femoral nail and the femoral nail made of a titanium alloy. Strains of the<br />

lag-screw and the surrounding bone are related to the proportional, that strain of the<br />

surrounding bone is greater the larger the strain of the lag-screw has been confirmed.<br />

Occurrence of strain in the direction of long axis by the buffer was placed in the center<br />

of each lag-screw, the promotion of bone healing was considered mechanism can be<br />

expected. Our results are not necessarily appropriate than Adaptive Remodeling 1) . it is<br />

the agenda that Some challenges the optimization of the shape and material in order to<br />

converge to the strain conditions a more appropriate.<br />

Fig 5. Analysis results of femoral nail (Strain).<br />

Fig 6. Analysis results of bone model (Strain).<br />

The stem of the original PEEK femoral nail is a feature that has been designed coneshaped<br />

junction of each parts in contact with the surface. the design has been realized<br />

continuous variation of material properties. Also in the stress distribution, stress that has<br />

been reduced in stages according to the distal of the stem had been confirmed (Figure 7).<br />

When attention is paid to the distal end of the stem, equivalent stress of the femoral nail<br />

made of a titanium alloy was average 1,760MPa, that of the original PEEK femoral nail<br />

was average 104MPa, stress concentration in the distal end of the stem can be relaxed<br />

has been confirmed at the original PEEK femoral nail (Figure 7). Because the stem was<br />

configured differently to a high elastic modulus from a low elastic modulus to the<br />

elastic modulus progressively, worked with the action is considered to alleviate the stem<br />

by elastic deformation of the distal end of the stem. Therefore, the original PEEK<br />

femoral nail was the effect of suppressing the fracture of the distal end of the stem was<br />

expected.