27.12.2012 Views

ARUP; ISBN: 978-0-9562121-5-3 - CMBBE 2012 - Cardiff University

ARUP; ISBN: 978-0-9562121-5-3 - CMBBE 2012 - Cardiff University

ARUP; ISBN: 978-0-9562121-5-3 - CMBBE 2012 - Cardiff University

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

3.3 Computational study<br />

The 3D model was meshed with unstructured tetrahedral grid and a sensitivity analysis<br />

was conducted doubling the number of elements and ensuring that pressure drop across<br />

the model did not change more than 2%. The mesh chosen for the in-silico study<br />

contained around 300,000 elements (Fig. 4, left). As in the experimental setup, the 3D<br />

model was coupled to the LPN comprising the UB, LB and P impedances merging into<br />

the constant pressure generator (16 mmHg), while excluding the additional compliance<br />

placed at the inlet. In fact, since the VAD flow signal was not measured, the inflow<br />

boundary condition (Qin) was assigned as the Fourier series of the velocity waveform<br />

derived from the sum of the three outflows (QUB, QLB, QP) recorded in-vitro. The C and<br />

R values, experimentally measured and extrapolated, respectively, were implemented in<br />

the LPN, resulting in a system of three ordinary differential equations and three<br />

algebraic equations. A multiscale coupling approach [8] was adopted imposing timevarying<br />

uniform pressures, calculated by the LPN, at each 3D outlet and flow rates,<br />

averaged over the sections, to the LPN. Blood was assumed as an incompressible<br />

Newtonian fluid with density of 1060 kg/m 3 and dynamic viscosity of 3.6·10 -3 Pa·s [12].<br />

A pulsatile simulation was run using commercial software (Fluent 12.1.4, ANSYS, Inc.,<br />

Canonsburg, PA, USA), with the implicit Euler method as the time integration<br />

technique for solving Navier-Stokes equations in the 3D domain and the explicit Euler<br />

method for solving the LPN system. Time step was fixed at 5·10 -5 s and four cycles<br />

were considered sufficient for stability of the solution. The average time to complete<br />

one cardiac cycle was about 24 hours, using an Intel® Core i7 (3GHz) personal<br />

computer. Pressure was monitored and averaged over three cross-sections of the 3D<br />

model, corresponding to the same locations of the pressure ports. All mean values were<br />

calculated over three cycles to compensate the variability of the experimental data.<br />

4. RESULTS AND DISCUSSION<br />

As expected, steady flow measurements performed on the valves indicated a non-linear<br />

behavior (∆P ≈ αQ 2 + βQ, being α and β constants, ∆P and Q the pressure drop and flow<br />

across the valve, respectively). Values of α and β coefficients for each non-linear<br />

resistance are displayed in Figure 2 in the fitting curve equations.<br />

Fig. 2 Non-linear pressure drop-flow<br />

relationships of the valves used in-vitro. The<br />

blue curve is relative to the upper and lower<br />

body resistances (RUB, RLB), the red curve to<br />

the pulmonary resistance (RP).<br />

Figure 3 shows the flow (left) and pressure<br />

(right) tracings recorded during the in-vitro<br />

and in-silico simulations. Overall,<br />

computational flows (dashed lines)<br />

satisfactorily matched the experimental<br />

data (solid lines), in terms of both<br />

waveforms and mean values. A maximum<br />

difference in mean flow of 17% was measured at the UB outlet (0.77 vs 0.90 L/min, invitro<br />

vs in-silico), where the computational model reported higher perfusion than in-

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!