did not allow gap closure. Maximum Von Mises stress and maximum equivalent strain<br />

on the proximal and distal osteotomy surfaces are shown in figures 4 and 5 respectively.<br />

Maximum stresses and strains within the osteotomy are found in the areas of contact at<br />

gap closure. These areas are found on the lateral side of the tibia, opposite the plate. For<br />

the entire construct, the highest stress found in any of the analyses is found on the head<br />

of the fifth screw of the Standard-Ti-Ti construct (Fig. 6). At 8162 MPa, the stresses on<br />

screw #5 are nearly 4 times the next highest amount.<br />

Figure 7 displays the Von Mises stress map of the plates and screws for each construct<br />

configuration with maximum stress positions labeled. The maximum stress for each<br />

construct is found on screw #5 except for the Reinforced-Pk-Pk construct, which has<br />

maximum stress on screw #1. Maximum stresses also occur on the near cortex for each<br />

construct except for the Standard-Ti-Ti construct upon which the maximum stress<br />

occurs at the base of the screw head near the bottom (lateral) side of the plate.<br />

5. DISCUSSION<br />

Reinforced-Pk-Ti Reinforced-Pk-Pk<br />

Max: 1470<br />

MPa<br />

Standard-Ti-Ti Standard-Pk-Pk<br />

Max: 8162<br />

MPa<br />

Max: 1014<br />

MPa<br />

Reinforcement Reinforcement<br />

Max: 1060<br />

MPa<br />

Figure 7: Front view of plate and screw stress results, 0 - 1000 MPa scale. Ti-reinforced PEEK plates<br />

also show stress results of inner reinforcement (front view) at bottom right of the figure.<br />

Interfragmentary motion (IFM) for the four constructs was high. Claes et al. found that<br />

0.5 mm IFM produces the most stable fracture callus for secondary fracture healing in<br />

fracture gaps of 2 mm or smaller. For larger gap sizes (3 to 6 mm), delayed healing can