Remote Operation<br />

Reference <strong>Power</strong> Related Commands<br />

PRFAsnn.nn<br />

PRFBsnn.nn<br />

Set Channel A Reference <strong>Power</strong> for graphic display. Sets 100% power reference<br />

point as a reference for automatic timing measurements. It also serves to center the<br />

pulse on the power axis so that displays and plots of pulses with low amplitudes are<br />

expanded to show more resolution.<br />

(s = polarity; nn.nn = dBm value)<br />

Set Channel B Reference <strong>Power</strong> for graphic display. Sets the 100% power reference<br />

point as a reference for automatic timing measurements. It also serves to center the<br />

pulse on the power axis so that displays and plots of pulses with low amplitudes are<br />

expanded to show more resolution.<br />

(s = polarity; nn.nn = dBm value)<br />

Min Value: -60.00 dBm, Max Value: +110.00 dBm<br />

Sample Program:<br />

10 P=0 ! CHANNEL A REFERENCE<br />

20 ! POWER IN DBM<br />

30 !<br />

40 ! CHANGE CHANNEL A REFERENCE<br />

50 ! POWER<br />

60 OUTPUT 704;PRFA;P<br />

70 !<br />

80 ! CHANGE CHANNEL B REFERENCE<br />

90 ! POWER<br />

100 OUTPUT 704;PRFB;P<br />

110 END<br />

☛ NOTE: RPOA and RPOB are operable only in the Graph Mode.<br />

RPOA<br />

Output Current Reference <strong>Power</strong> for Channel A. If the PPM is in the linear mode it<br />

will output:<br />

RFALnn.nnEsnn<br />

(nn.nn = Watts; Esnn = exponent)<br />

If the PPM is in the dBm mode it will output:<br />

RFADsnn.nn<br />

(nn.nn = dBm; s = polarity)<br />

RPOB<br />

Output Current Reference <strong>Power</strong> for Channel B. Linear mode output is:<br />

RFBLnn.nnEsnn<br />

(nn.nn = Watts; Esnn = exponent)<br />

The dBm mode output is:<br />

RFBDsnn.nn<br />

(nn.nn = dBm; s = polarity)<br />

<strong>Manual</strong> No. 20790, Rev C, November 1998 3-27<br />