<strong>Series</strong> <strong>8500A</strong> <strong>Peak</strong> <strong>Power</strong> <strong>Meter</strong>s<br />

5.2 Performance Testing<br />

The tests in this section are to determine that the performance of the Model 8501A or 8502A <strong>Peak</strong> <strong>Power</strong><br />

<strong>Meter</strong>s (PPM) is within the parameters of the specifications given for the instrument. Assemble all of the<br />

required test equipment and the PPM to be tested. Connect all equipment to the ac power line, turn it on,<br />

and allow at least 30 minutes for it to warm up before starting the test.<br />

5.2.1 Equipment Required<br />

The equipment listed in Table 5-1 is required for Performance Testing. Equivalent or similar test<br />

equipment can be substituted.<br />

Table 5-1. Required Test Equipment<br />

Description Representative Model Key Characteristics<br />

CW Thermistor <strong>Power</strong> <strong>Meter</strong> HP Model 432B Instrument accuracy of at least<br />

0.5%<br />

Thermistor Mount HP Model 8478B ≤1.1 SWR @ 1 GHz<br />

Digital Voltmeter (DVM) Fluke Model 8600A ±0.05% accuracy<br />

Directional Coupler Narda Model 3002, 10 dB ≤1.15 SWR @ 1 GHz<br />

Step Attenuator, 0 to 50 dB in<br />

10 dB increments<br />

Weinschel Model 1<br />

≤1.15 SWR @ 1 GHz Accuracy<br />

+20 dBm @ 1 GHz & pulse mod<br />

Low Pass Filter HP Model 360B >30 dB attenuation @ 2 GHz<br />

Precision Pulse Generator Wavetek Model 178 10 ppm Frequency Accuracy<br />

Return Loss Bridge Wiltron 63N50 or 97NF50 Directivity 36 dB. Freq. Range<br />

from 1 GHz to max. freq. of<br />

detector in use<br />

Scalar Network Analyzer <strong>Giga</strong>-<strong>tronics</strong> Model 8003 Frequency Range as above<br />

Precision DC Source Datel Model DVC <strong>8500A</strong> 0.05% Accuracy<br />

Graphics Plotter HP Colorpro Model 7440A GPIB compatible<br />

5-2 <strong>Manual</strong> No. 20790, Rev C, November 1998<br />

Superceded by Revision D, March 2009