Operation<br />

2.7.2 Self-Calibration Failures<br />

In the event of a malfunction such as a burned out detector diode or if you forget to connect the detector<br />

to the calibrator, the Self-Cal function will fail. If this occurs, the following error menu will be displayed:<br />

A Calibration Error Exists<br />

Press [CLEAR] to continue<br />

Press any Units key to repeat cal<br />

Press [CLEAR] to return the instrument to the last mode of operation.<br />

Selecting any Units key gives you the opportunity to correct a simple error such as not connecting the<br />

detector to the calibrator. This can be accomplished, and then the Self-Cal procedure can be repeated.<br />

See Input/Output & Calibration Failure descriptions in Section 6.3.2 for the procedure to follow if the<br />

PPM fails the Self-Cal.<br />

2.7.3 Auto-Zero Function<br />

To initiate the Auto-Zero function, turn off the RF power to the detector to be zeroed and then press<br />

[MENU] [F3] to get to the Self-Cal/Auto-Zero menu. Then press [F2] to Auto-Zero and follow the<br />

displayed prompts.<br />

If the Model 8502A is in use, channel A must be auto-zeroed before channel B. If desired, channel A can<br />

be bypassed by pressing [CLEAR] and going directly to channel B.<br />

<strong>Manual</strong> No. 20790, Rev C, November 1998 2-15<br />

Superceded by Revision D, March 2009