7. Press [F1].<br />

8. The DVM should read -5.000 V ±35 mV.<br />

9. If the unit being tested is an 8502A, press [F2] and repeat Steps 3 through 8 after moving the<br />

rear cable to the ANALOG B output.<br />

E. Thermistor Sense Circuit.<br />

1. Press [CLEAR] [4] [F1]<br />

2. The reading should be approximately 5000 mV at a detector temperature of 30 °C (82.4 °F). The<br />

reading will vary about ±0.2 V/°C with any increase or decrease in temperature from the 30 °C<br />

level (a new reading is taken each time F1 is pressed.)<br />

3. Press [CLEAR] [CLEAR].<br />