Operation<br />

2.2.1 PCB-Oriented Voltage Selector and Fuse Holder<br />

The number on the selector card, visible through the window, indicates the nominal line voltage to which<br />

the power meter must be connected.<br />

The line voltages and fuse ratings are:<br />

Line Voltage<br />

Fuse Rating<br />

100/120 Vac, ±10%, 50, 60 or 400 Hz ±5% 1.0 AMP Slo-Blo<br />

220/240 Vac, ±10%, 50, 60 or 400 Hz ±5% 2.0 AMP Slo-Blo<br />

To select a different operating line voltage and fuse, refer to Figure 2-2 and proceed as follows:<br />

1. Open the cover door, rotate the fuse-pull to the left, and remove the fuse.<br />

2. Select the operating voltage by orienting the PC board so that the correct voltage label is on the<br />

top left side.<br />

3. Push the board firmly back into the module slot.<br />

4. Rotate the fuse-pull back into the normal position and reinsert the fuse into the holder. Use care<br />

to select the correct fuse value.<br />

Operating voltage is shown<br />

in the module window<br />

Figure 2-2. PCB Voltage Selector and Fuse Holder<br />

<strong>Manual</strong> No. 20790, Rev C, November 1998 2-3<br />

Superceded by Revision D, March 2009