Manual - 8500A Series Peak Power Meter - Giga-tronics
Manual - 8500A Series Peak Power Meter - Giga-tronics
Manual - 8500A Series Peak Power Meter - Giga-tronics
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
Testing and Calibration<br />
5.2.7 Analog Output Accuracy Test<br />
Set the RF Source to generate 0 dBm CW and connect the channel A detector to the output. Connect the<br />
DVM to the channel A analog output on the rear of the PPM and, using the procedure in step 1 of the<br />
<strong>Power</strong> Linearity Test, set the PPM to read CW with an averaging factor of 900. Perform the following<br />
steps:<br />
1. Adjust the RF Source so that the PPM reads exactly 0.00 dBm. Record the reading of the DVM.<br />
It should be less than 10 mV.<br />
2. Adjust the RF Source so that the PPM reads exactly -10.00 dBm. Subtract the DVM reading<br />
recorded in Step 1 from the current reading. The magnitude of the difference should be<br />
1.000V ±5 mV.<br />
3. If a Model 8502A is being tested, repeat Steps 1 and 2 after moving the DVM to the channel B<br />
analog output.<br />
5.2.8 Voltage Proportional to Frequency Test<br />
Press [MENU] (3) [F3] to choose the source of frequency correction. Then press [F2] to select the rear<br />
panel input. The input characteristics will then be specified as follows:<br />
Voltage/GHz = 1.00<br />
Sweeper Start Voltage = 0.000<br />
Sweeper Start Frequency = 0.000<br />
Adjust the Precision DC Source to 0.00 V and connect it to the frequency input on the rear panel of the<br />
PPM. Select the CW Mode display. The FREQ readout at the bottom of the display should read<br />
0.000 GHz. Set the Precision DC Source to exactly 10.00 V. The display should read<br />
10.000 GHz ±0.050 GHz.<br />
5.2.9 Detector Return Loss Test<br />
Use the 8003 or another return loss test setup appropriate to the frequency range of the detectors to be<br />
used with the PPM. Measure the return loss of the detector(s) used with the PPM. Specifications depend<br />
on the type of detector, its frequency range, and the type of connector being used. Refer to Appendix D<br />
to determine the specifications for the particular detector under test. When using this equipment, be sure<br />
to allow for the bridge directivity contribution to measurement uncertainty (in dB); this must be<br />
subtracted from the value of return loss specified for the detector.<br />
5.2.10 Plotter Output/IEEE-488 Interface<br />
Connect the signal generator to the PPM and display a pulse waveform in the Graph Mode. The pulse<br />
parameters, frequency, etc. are arbitrary for the purposes of testing the GPIB plot output of the PPM.<br />
1. Use an IEEE-488 cable to connect the graphics plotter to the GPIB output connection on the rear<br />
panel of the PPM. Note the plotter address as selected on the plotter.<br />
2. Press [MENU] (6) [F2] to enter the plotter address. If the current address is different from the<br />
desired address, enter the correct address number.<br />
3. Press [MENU] (2) [F3] to begin the plot.<br />
This completes the Performance Verification Tests of the PPM. If the instrument performed as given in<br />
this procedure, it is correctly calibrated and within published specifications.<br />
<strong>Manual</strong> No. 20790, Rev C, November 1998 5-7<br />
Superceded by Revision D, March 2009