Remote Operation<br />

When plots are output from the PPM, the pulse profile will always be in a linear format including when<br />

the reference power is labeled in dBm.<br />

3.3.11 Programming Notes<br />

The programming commands for the PPM can be strung together as long as the maximum length<br />

limitations specified in the General Information discussion are not exceeded.<br />

This feature enables procedures to be formatted in a logical manner, and reduces the overall number of<br />

programming lines required. The following sample program shows how this can be done.<br />

10 CLEAR 704! DEVICE CLEAR<br />

20 OUTPUT 704;ASDL200,AVAS25,AUTO<br />

30 OUTPUT 704;PLCNabcdefg;PLPNqwerty;<br />

40 RSFL,PLTM<br />

50 SEND 7;UNL<br />

60 SEND 7;LISTEN 6<br />

70 SEND 7;TALK 4<br />

80 SEND 7;DATA<br />

90 WAIT 180<br />

100 SEND 7;UNT UNL<br />

110 END<br />

Line 20: Sets the maximum pulse width for autoscaling to 200 microseconds; sets the averaging for<br />

autoscaling to 25 averages; instructs the PPM to autoscale.<br />

Lines 30 & 40: Set the code numbers for the plot to abcdefg; sets the part number for the plot to qwerty;<br />

instructs the PPM to find the rise and fall times of the displayed pulse; instructs the PPM to place the<br />

markers as defined and to plot the displayed pulse profile.<br />

The remaining lines in the sample program are used for bus handling.<br />

Care should be exercised when grouping commands. If conflicting commands are sent in the same string,<br />

the last one to be received will be implemented. For example:<br />

OUTPUT 704;MRKA,CDLA5,PKPA<br />

This program line would result in the cursor delay being set at 5 µs, but the PPM would enter the<br />

channel A <strong>Peak</strong> Mode, not the Marker Mode as indicated by the MRKA portion. In other words, the PPM<br />

does not execute any of the commands until (CR) (LF) are received at the end of the program line.<br />

<strong>Manual</strong> No. 20790, Rev C, November 1998 3-35<br />

Superceded by Revision D, March 2009