Errata Sheet<br />

for<br />

<strong>Series</strong> <strong>8500A</strong> <strong>Peak</strong> <strong>Power</strong> <strong>Meter</strong><br />

Operation & Maintenance <strong>Manual</strong><br />

<strong>Manual</strong> P/N 20790<br />

Issued March 10, 2009<br />

Correction: Change in Performance Specifications<br />

Description of Correction: This change applies to the <strong>Giga</strong>-<strong>tronics</strong> <strong>8500A</strong> series <strong>Peak</strong> <strong>Power</strong> <strong>Meter</strong> with<br />

the model 17071A 40 GHz <strong>Power</strong> Sensor. The change is to the measurement <strong>Power</strong> Range only:<br />

Old 17071A specification:<br />

750 MHz to 40 GHz<br />

<strong>Peak</strong> <strong>Power</strong> Range: -20 to +20 dBm and CW <strong>Power</strong> Range: -40 to +20 dBm<br />

New 17071A specification:<br />

750 MHz to 18 GHz<br />

<strong>Peak</strong> <strong>Power</strong> Range: -20 to +20 dBm and CW <strong>Power</strong> Range: -40 to +20 dBm<br />

>18 GHz to 40 GHz<br />

<strong>Peak</strong> <strong>Power</strong> Range: -20 to +0 dBm and CW <strong>Power</strong> Range: -40 to +0 dBm<br />

All other specifications remain the same.<br />

© 2009 <strong>Giga</strong>-<strong>tronics</strong> Incorporated.<br />

Superceded by Revision D, March 2009