<strong>Series</strong> <strong>8500A</strong> <strong>Peak</strong> <strong>Power</strong> <strong>Meter</strong>s<br />

5.3 Calibration Procedures<br />

Using all solid state components, the Model 8501A/8502A meters are extremely rugged and reliable.<br />

Consequently, there is very little drift due to component aging, and adjustments to the units are rarely<br />

required. If measurements indicate that an adjustment is set within the indicated range, do not attempt to<br />

make it exact. It is often the case that variations in the equipment being used to make the test account for<br />

small differences in measured values. Since some adjustments can be interactive, be absolutely sure that<br />

an adjustment is required before making it.<br />

These procedures should be performed whenever the instrument is due for routine calibration, or at any<br />

time calibration is required due to environmental or physical conditions. All test points and adjusting<br />

components are labeled and are accessible at the top of the PC boards. It is not necessary to remove any<br />

PC boards from the unit to perform these calibration procedures.<br />

Figure 5-3 is a reference locater for the various components required for adjusting purposes during the<br />

calibration procedure.<br />

If you need to calibrate any of the RF detectors used with the PPM (such as after replacing the diode<br />

element), you can do it in the field with the <strong>Giga</strong>-<strong>tronics</strong> PROM Programming Accessory Kit, P/N 16976.<br />

This kit contains all of the necessary components and instructions required for proper detector<br />

calibration. If the kit is not available, the detector can be returned to the factory for calibration.<br />

5-8 <strong>Manual</strong> No. 20790, Rev C, November 1998<br />

Superceded by Revision D, March 2009