<strong>Series</strong> <strong>8500A</strong> <strong>Peak</strong> <strong>Power</strong> <strong>Meter</strong>s<br />

1.3 Default Settings<br />

Factory-programmed settings of various user definable functions have the following defaults whenever<br />

there have been no changes made in the settings by the operator during any testing routines.<br />

Function<br />

Averaging<br />

Internal Trigger Level (A & B)<br />

Frequency<br />

Offset<br />

Cursor Delay (A & B)<br />

Start Delay<br />

Delay Window (A & B)<br />

Reference <strong>Power</strong> Level (A & B)<br />

Reference Delay (A & B)<br />

Cal Factor<br />

Mode<br />

Trigger Mode<br />

Default Setting<br />

CW: 4 samples; <strong>Peak</strong>: 4 samples<br />

-10 dBm<br />

1.00 GHz<br />

0 dB<br />

1.0000 microseconds<br />

0.0000 nanoseconds<br />

10.0000 microseconds<br />

0 dBm<br />

0.0000 microseconds<br />

0 dB<br />

Channel A, CW power, dBm<br />

Channel A, Internal<br />

Pulse Width Start 50%<br />

Pulse Width End 50%<br />

Rise Time Start 10%<br />

Rise Time End 90%<br />

Fall Time Start 90%<br />

Fall Time End 10%<br />

Marker 1 2.5% (A and B)<br />

Marker 2 17.8% (A and B)<br />

Marker 3 46.9% (A and B)<br />

Marker 4 90% (A and B)<br />

The following settings are not affected by power-on, initialization, and setup store or recall:<br />

Maximum <strong>Power</strong><br />

Minimum <strong>Power</strong><br />

+20.5 dBm<br />

-45.0 dBm<br />

Autoscale Average Number 4<br />

Autoscale Initial Delay 75 µs<br />

Plotter Address 6<br />

PPM Address 4<br />

Source of Frequency Correction<br />

Frequency Display<br />

a) User supplied (as input, default = 1GHz)<br />

b) VpropF - parameters not subject to store/<br />

recall or initialization<br />

c) Cal Factor - not subject to store/recall or<br />

initialization<br />

Displayed<br />

1-8 <strong>Manual</strong> 20790, Rev. C, November 1998<br />