13.11.2014 Views

Manual - 8500A Series Peak Power Meter - Giga-tronics

Manual - 8500A Series Peak Power Meter - Giga-tronics

Manual - 8500A Series Peak Power Meter - Giga-tronics

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

Operation<br />

2.11 High <strong>Power</strong> Measurements<br />

One of the most important things to remember when using the PPM is that it employs balanced, zero<br />

bias, Schottky diode detectors as power detectors. Due to this highly sensitive, highly accurate<br />

configuration, the maximum power that the detectors can handle before burning out is +23 dBm<br />

(200 mW). This limitation is regardless of duty cycle.<br />

2.11.1 <strong>Power</strong> Warning - Max/Min <strong>Power</strong> Limits<br />

The PPM has a built-in warning system to advise you of mild overloads to the detectors. Large overloads<br />

which are usually the result of a failure in the device under test or leaving an attenuator out of the test<br />

setup, cannot be prevented because of the speed at which these failures happen. However, you will be<br />

advised by a prompt when the power level exceeds the preset power warning level.<br />

The default value for the MAX <strong>Power</strong> warning point is +20.5 dBm. You can set this to anywhere between<br />

-10 and +21 dBm.<br />

The PPM also has a lower level limit (MIN). This is mainly used to avoid displaying power readings<br />

outside the lower power range of the instrument. When the power level being measured by the PPM is<br />

below the lower level limit, the readout will display a series of dashes to indicate that the power being<br />

measured is too low for the PPM to take an accurate measurement.<br />

The default value for the MIN <strong>Power</strong> warning point is -45.0 dBm. You can set this anywhere between -15<br />

and -50 dBm. To change the MAX and MIN power limits, press [MENU] (7) [F1] [mm.mm] [dBm]<br />

[nn.nn] [dBm], where mm.mm is the maximum power limit, and nn.nn is the minimum power limit.<br />

2.12 High <strong>Power</strong> Measurement Procedures<br />

The procedure for making absolute CW or <strong>Peak</strong> power measurements under high power conditions is as<br />

follows:<br />

1. Approximate the <strong>Peak</strong> power of the signal to be measured.<br />

2. After 30 minutes warm-up, Self-Cal the PPM detectors.<br />

3. Choose an appropriate attenuator or coupler to be used in the test setup to reduce the power<br />

being measured to a safe level for the detectors. The power at the detector input must be less<br />

than +20 dBm.<br />

4. When making this approximation, be sure to use worst case power levels. If in doubt, a 10 dB<br />

attenuator can be used directly on the detector input in addition to the selected high power<br />

coupler or attenuator.<br />

5. Ensure that the RF power of the source to be measured is turned OFF.<br />

6. Connect the high power coupler to the RF source to be measured as shown in Figure 2-14. If a<br />

high power attenuator is used, connect as shown in Figure 2-15.<br />

7. Connect the PPM detector with its attenuator to the coupled output of the high power coupler<br />

(Figure 2-14), or to the output of the high power attenuator (Figure 2-15).<br />

<strong>Manual</strong> No. 20790, Rev C, November 1998 2-37<br />

Superceded by Revision D, March 2009

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!