Theory of Operation<br />

A5 Test Points<br />

The test points in Table 4-4 can be checked for the correct signals on the A5 Delay Line board:<br />

Table 4-4. A5 Test Points<br />

Test Point<br />

TP1<br />

TP2<br />

TP3<br />

TP4<br />

TP5<br />

TP6<br />

TP7<br />

TP8<br />

TP9<br />

TP10<br />

TP11<br />

TP12<br />

TP13<br />

TP14<br />

TP15<br />

TP16<br />

TP17<br />

TP18<br />

Description<br />

Output of U29A. Trigger signal that tells when the board has been<br />

triggered and reset. Can trigger a scope when testing the board.<br />

Signal goes low at trigger and high at sample time.<br />

-5.2 V<br />

5 V Common<br />

+5 V<br />

Clock Frequency (C79 adjusts)<br />

Output of U36, U37 DAC, which determines the 0 to 25.5 ns<br />

threshold.<br />

U34 comparator input<br />

U34 comparator input<br />

U37B hold amplifier output<br />

10 V Reference<br />

U34 comparator output<br />

5 V Common<br />

U39 comparator output (trigger signal)<br />

U39 comparator output (trigger signal)<br />

U40. Trigger level DAC<br />

+15 V<br />

15 V Common<br />

-15 V<br />

<strong>Manual</strong> No. 20790, Rev C, November 1998 4-19<br />