13.11.2014 Views

Manual - 8500A Series Peak Power Meter - Giga-tronics

Manual - 8500A Series Peak Power Meter - Giga-tronics

Manual - 8500A Series Peak Power Meter - Giga-tronics

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

Remote Operation<br />

180 ! COMMANDS WILL INITIATE THE<br />

190 ! PPM’S PLOT ROUTINE. THE<br />

200 ! ADDRESS OF THE PLOTTER AND<br />

210 ! PPM MUST BE SET CORRECTLY -<br />

220 ! IN THIS CASE THE PPM IS AT 704,<br />

230 ! AND THE PLOTTER IS AT 706.<br />

240 OUTPUT 704;PLCN&C$<br />

250 ! SEND CODE NUMBER TO PPM<br />

260 OUTPUT 704;PLPN&P$<br />

270 ! SEND PART NUMBER TO PPM<br />

280<br />

290 ! THESE ARE THE COMMANDS TO<br />

300 ! INITIATE THE PLOT ROUTINE:<br />

310<br />

320 OUTPUT 704;PLOT<br />

330 ! (PUTS PPM IN PLOT OUTPUT MODE)<br />

340<br />

350 SEND 7;UNL<br />

360 SEND 7;LISTEN 6<br />

370 ! (LISTEN ADDRESSES THE PLOTTER)<br />

380 SEND 7;TALK 4<br />

390 ! (TALK ADDRESSES THE PPM)<br />

400<br />

410 SEND 7;DATA<br />

420 ! (RELEASES ATN LINE ALLOWING<br />

430 ! PLOTTING TO BEGIN)<br />

440<br />

450 WAIT 180<br />

460 ! WAIT UNTIL PLOTTING HAS<br />

470 ! FINISHED BEFORE UNTALKING<br />

480 ! THE PPM. PLOT TAKES ABOUT<br />

490 ! 3 MINUTES<br />

500<br />

510 ! UNTALK AND UNLISTEN ALL BUS<br />

520 ! DEVICES<br />

530 SEND 7;UNT UNL<br />

540 END<br />

<strong>Manual</strong> No. 20790, Rev C, November 1998 3-9<br />

Superceded by Revision D, March 2009