Remote Operation<br />

3.6 Status Code Decimal Values<br />

The decimal values of the various status codes minus the 7th bit are as listed below.<br />

To determine the value with the 7th bit high, add 64 to the values listed in Table 3-1.<br />

Table 3-1. Status Code Values and Conditions<br />

Decimal<br />

Value<br />

Condition<br />

0 Operation normal, no condition to report<br />

1 Output ready to be sent over GPIB<br />

10 Self test passed<br />

20 Calibration passed<br />

25 Autozero passed<br />

27 Autoscaling completed and passed<br />

31 If RISE, PLWD, FALL, RSFL, RSWD, WDFL, PUFT, PURT, OR PUWD all measurements<br />

successfully made<br />

33 Command string syntax error, command(s) not processed<br />

34 Command string has invalid argument, command(s) not processed<br />

35 Command string(s) incomplete when unlistened, string(s) sent so far are lost, command(s) not<br />

processed<br />

36 PPM command buffer overflow. Limit of 128 character string exceeded. String lost, no<br />

commands processed<br />

37 Cannot execute MKDF, markers to be measured are undefined. No command(s) processed<br />

38 Command not executed, PPM not in REMOTE state<br />

39 Command not implemented, no commands processed<br />

40 GET sent when not in REMOTE<br />

41 GET or UPDN command sent when not in the Update Data Trigger Reset Output Mode. If<br />

UPDN command, no command(s) processed<br />

43 Not in Graph Mode or Marker sub-mode, must be in Graph Mode for AUTO (Autoscale<br />

Graph), or PLOT commands to function. Must be in Marker sub-mode for marker commands<br />

to function. No command(s) processed<br />

44 Autoscaling, PLOT, DUMG, pulse parameter, or a marker placement command has been<br />

specified in the command string and a mode change requested later at some point in the same<br />

command string. Mode change not processed<br />

45 STAT command sent while SRQ enabled, STAT command not active<br />

46 SRQ buffer overflow, SRQ enabled and not being read quickly enough by the controller<br />

48 Cannot enter Fast Measurement Mode. Either in linear, graph, or marker mode. No<br />

command(s) processed<br />

49 Cannot execute RECL or WATT commands, or any graph or marker commands in the Fast<br />

Measurement Mode. No command(s) processed<br />

50 Error in reading PROM data from detector A<br />

51 Error in reading PROM data from detector B<br />

<strong>Manual</strong> No. 20790, Rev C, November 1998 3-41<br />

Superceded by Revision D, March 2009