Remote Operation<br />
3.6 Status Code Decimal Values<br />
The decimal values of the various status codes minus the 7th bit are as listed below.<br />
To determine the value with the 7th bit high, add 64 to the values listed in Table 3-1.<br />
Table 3-1. Status Code Values and Conditions<br />
Decimal<br />
Value<br />
Condition<br />
0 Operation normal, no condition to report<br />
1 Output ready to be sent over GPIB<br />
10 Self test passed<br />
20 Calibration passed<br />
25 Autozero passed<br />
27 Autoscaling completed and passed<br />
31 If RISE, PLWD, FALL, RSFL, RSWD, WDFL, PUFT, PURT, OR PUWD all measurements<br />
successfully made<br />
33 Command string syntax error, command(s) not processed<br />
34 Command string has invalid argument, command(s) not processed<br />
35 Command string(s) incomplete when unlistened, string(s) sent so far are lost, command(s) not<br />
processed<br />
36 PPM command buffer overflow. Limit of 128 character string exceeded. String lost, no<br />
commands processed<br />
37 Cannot execute MKDF, markers to be measured are undefined. No command(s) processed<br />
38 Command not executed, PPM not in REMOTE state<br />
39 Command not implemented, no commands processed<br />
40 GET sent when not in REMOTE<br />
41 GET or UPDN command sent when not in the Update Data Trigger Reset Output Mode. If<br />
UPDN command, no command(s) processed<br />
43 Not in Graph Mode or Marker sub-mode, must be in Graph Mode for AUTO (Autoscale<br />
Graph), or PLOT commands to function. Must be in Marker sub-mode for marker commands<br />
to function. No command(s) processed<br />
44 Autoscaling, PLOT, DUMG, pulse parameter, or a marker placement command has been<br />
specified in the command string and a mode change requested later at some point in the same<br />
command string. Mode change not processed<br />
45 STAT command sent while SRQ enabled, STAT command not active<br />
46 SRQ buffer overflow, SRQ enabled and not being read quickly enough by the controller<br />
48 Cannot enter Fast Measurement Mode. Either in linear, graph, or marker mode. No<br />
command(s) processed<br />
49 Cannot execute RECL or WATT commands, or any graph or marker commands in the Fast<br />
Measurement Mode. No command(s) processed<br />
50 Error in reading PROM data from detector A<br />
51 Error in reading PROM data from detector B<br />
<strong>Manual</strong> No. 20790, Rev C, November 1998 3-41<br />