13.11.2014 Views

Manual - 8500A Series Peak Power Meter - Giga-tronics

Manual - 8500A Series Peak Power Meter - Giga-tronics

Manual - 8500A Series Peak Power Meter - Giga-tronics

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

Operation<br />

2.10.5 <strong>Power</strong> Ratio Measurements<br />

To make ratio measurements, the power measured by both channels must be within the specified power<br />

measurement range of the PPM system. When operating in CW, this power range is -40 to +20 dBm. In<br />

the <strong>Peak</strong> Mode, this power range is -20 to +20 dBm. <strong>Power</strong> readings taken in the Ratio Mode outside of<br />

these ranges will result in an error display.<br />

The Ratio Mode only provides the ratio of A/B, not vice versa. The PPM can ratio two CW signals, two<br />

<strong>Peak</strong> signals, or a combination of both. The following procedures show how to perform each of these<br />

ratio’ing operations.<br />

Ratio’ing Two CW Signals<br />

This is the easiest ratio function of the PPM. Proceed as follows:<br />

1. Allow the PPM to warm up for at least 30 minutes. Then Self-Calibrate both channels. From<br />

power-on, press [CW] [MENU] (1) [F3] [F1].<br />

Follow the displayed prompts to complete the Self-Cal. After Self-Cal has completed, the PPM<br />

will be measuring CW power in channel A.<br />

2. Select the Ratio Display Mode by pressing [MENU] (7) [F2].<br />

3. The default mode for power ratio measurements is for <strong>Peak</strong> power on both channels. To set both<br />

channels for CW operation, press [A] [F2] [B] [F2].<br />

The PPM will now be reading the ratio of A/B. Normally this is expressed in dB, however, it can also be<br />

expressed as a linear reading (such as 19654 EXP=-4) by pressing [dB/mW]. The display indicates that<br />

two CW signals are being ratio’ed by displaying CW/CW in the lower right hand corner of the display.<br />

Remember that the frequency of operation or Cal Factor must be entered using the FREQ key, and any<br />

desired Offsets should also be included. These two functions are described in Sections 2.8.3 and 2.8.4.<br />

Ratio’ing Two <strong>Peak</strong> Signals<br />

The point in time at which readings of <strong>Peak</strong> power are taken in both channels for ratio’ing purposes is<br />

determined by the setting of the Cursor Delay. Although it is most desirable that both pulses being<br />

monitored occur at exactly the same time, in practice this would be a very rare event. Therefore, the<br />

signals to be measured must be carefully evaluated to ensure that accurate readings are taken.<br />

The Cursor Delay must be set so that samples are being taken on both the channel A and B pulses. See<br />

Figures 2-12 and 2-13 for an examples of good and bad Cursor Delay settings in this situation.<br />

The following are guidelines that can be used when ratio’ing two <strong>Peak</strong> signals:<br />

1. Make a single channel <strong>Peak</strong> measurement using the Graph Mode (Autoscaling is recommended).<br />

2. Set the Cursor Delay so that the power reading (CSR PWR) is being taken close to the center of<br />

the pulse (see Figure 2-12).<br />

Figure 2-12. Typical <strong>Peak</strong> <strong>Power</strong> Ratio Display<br />

<strong>Manual</strong> No. 20790, Rev C, November 1998 2-35<br />

Superceded by Revision D, March 2009

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!