13.11.2014 Views

Manual - 8500A Series Peak Power Meter - Giga-tronics

Manual - 8500A Series Peak Power Meter - Giga-tronics

Manual - 8500A Series Peak Power Meter - Giga-tronics

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

<strong>Series</strong> <strong>8500A</strong> <strong>Peak</strong> <strong>Power</strong> <strong>Meter</strong>s<br />

Auxiliary Inputs/Outputs<br />

Monitor:<br />

Analog Output Coefficient:<br />

Offset:<br />

Trigger Input:<br />

RF Blanking:<br />

V PROP F:<br />

Provides a voltage proportional to the detected RF envelope.<br />

Rise time is typically 20 ns, output impedance is nominally 50<br />

ohms.<br />

100 mV/dB ±0.5 mV (0 V = 0 dBm)<br />

±10 mV<br />

TTL<br />

TTL open collector low during zeroing. Used to control power<br />

source.<br />

(V/GHz) Allows direct entry of frequency from RF power sources<br />

equipped with a VpropF output.<br />

GPIB Interface: Per IEEE STD 488-1978<br />

GPIB Indicators:<br />

REM, TLK, LSN, SRQ, LLO<br />

Remote Operation<br />

Direct Plot Output:<br />

GPIB Address:<br />

IEEE Interface Functions:<br />

Complete setup and measurement capabilities accessible via<br />

GPIB (IEEE 488). Reporting of errors, malfunctions, operational<br />

status, and self-test diagnostics available through serial poll<br />

capability.<br />

Outputs hardcopy pulse profile plots, with time, date and part<br />

identification, to a GPIB plotter.<br />

Selectable from front panel<br />

SH1, AH1, T6, L4, SR1, RL1, PP0, DC1, DT1, TE0, LE0<br />

General Specifications<br />

Stored Setups:<br />

<strong>Power</strong>-On Self-Test (POST):<br />

Reset Control<br />

(rear panel):<br />

Saves settings at power down, and has 10 additional setups in<br />

non-volatile memory.<br />

POST is optionally performed at any time. A diagnostic code<br />

indicates the cause and location of any errors.<br />

Returns the instrument to preset (default) condition.<br />

Design and Construction: To the intent of MIL-T-28800C, Type III, Class 5, Style E or F,<br />

Color R<br />

<strong>Power</strong> Requirements: 100, 120, 220, or 240 Vac ±10%, 50, 60, or 400 Hz ±5%<br />

<strong>Power</strong> Consumption:<br />

Operating Temperature:<br />

Non-operating Temperature:<br />

Approximately 100 VA<br />

0 to 50 °C (32 to 122 °F)<br />

-40 to +65 °C (-40 to +149 °F)<br />

1-6 <strong>Manual</strong> 20790, Rev. C, November 1998<br />