<strong>Series</strong> <strong>8500A</strong> <strong>Peak</strong> <strong>Power</strong> <strong>Meter</strong>s<br />
Auxiliary Inputs/Outputs<br />
Monitor:<br />
Analog Output Coefficient:<br />
Offset:<br />
Trigger Input:<br />
RF Blanking:<br />
V PROP F:<br />
Provides a voltage proportional to the detected RF envelope.<br />
Rise time is typically 20 ns, output impedance is nominally 50<br />
ohms.<br />
100 mV/dB ±0.5 mV (0 V = 0 dBm)<br />
±10 mV<br />
TTL<br />
TTL open collector low during zeroing. Used to control power<br />
source.<br />
(V/GHz) Allows direct entry of frequency from RF power sources<br />
equipped with a VpropF output.<br />
GPIB Interface: Per IEEE STD 488-1978<br />
GPIB Indicators:<br />
REM, TLK, LSN, SRQ, LLO<br />
Remote Operation<br />
Direct Plot Output:<br />
GPIB Address:<br />
IEEE Interface Functions:<br />
Complete setup and measurement capabilities accessible via<br />
GPIB (IEEE 488). Reporting of errors, malfunctions, operational<br />
status, and self-test diagnostics available through serial poll<br />
capability.<br />
Outputs hardcopy pulse profile plots, with time, date and part<br />
identification, to a GPIB plotter.<br />
Selectable from front panel<br />
SH1, AH1, T6, L4, SR1, RL1, PP0, DC1, DT1, TE0, LE0<br />
General Specifications<br />
Stored Setups:<br />
<strong>Power</strong>-On Self-Test (POST):<br />
Reset Control<br />
(rear panel):<br />
Saves settings at power down, and has 10 additional setups in<br />
non-volatile memory.<br />
POST is optionally performed at any time. A diagnostic code<br />
indicates the cause and location of any errors.<br />
Returns the instrument to preset (default) condition.<br />
Design and Construction: To the intent of MIL-T-28800C, Type III, Class 5, Style E or F,<br />
Color R<br />
<strong>Power</strong> Requirements: 100, 120, 220, or 240 Vac ±10%, 50, 60, or 400 Hz ±5%<br />
<strong>Power</strong> Consumption:<br />
Operating Temperature:<br />
Non-operating Temperature:<br />
Approximately 100 VA<br />
0 to 50 °C (32 to 122 °F)<br />
-40 to +65 °C (-40 to +149 °F)<br />
1-6 <strong>Manual</strong> 20790, Rev. C, November 1998<br />
Superceded by Revision D, March 2009