<strong>Series</strong> <strong>8500A</strong> <strong>Peak</strong> <strong>Power</strong> <strong>Meter</strong>s<br />

5.3.4 Initialization<br />

A. Turn the instrument on.<br />

The PPM will automatically perform a self-test. At the completion of the self-test, the PPM will<br />

return to the last setup before power down. If the self-test failed, note the failure number and proceed<br />

by pressing [CLEAR].<br />

B. Calibrate the Detector.<br />

The following portion of the Calibrator procedure needs to be done only if the detector being used is<br />

different than the one that was last used with the PPM.<br />

1. Press [CLEAR].<br />

2. Connect the detector to the Calibrator output (use channel A detector if the unit is an 8502A).<br />

3. Press [UNITS] and wait for the calibration cycle to complete.<br />

4. Disconnect the detector from the Calibrator output.<br />

5. If the unit is an 8502A and a channel B detector is connected, perform the following steps.<br />

a. Connect the channel B detector to the Calibrator output.<br />

b. Press [CLEAR] [UNITS] and wait for the calibration cycle to complete.<br />

c. Disconnect the detector from the Calibrator output.<br />

C. Set the Internal Clock<br />

The Internal Clock can be set as follows:<br />

1. Press [MENU] (5) [F3] [F1].<br />

2. Enter the correct date ([mmddyy] [UNITS]).<br />

3. Enter the correct time ([hhmmss] [UNITS] [F3]).<br />

5.3.5 A8 Front Panel Interface Board<br />

A. Display Brightness Adjust<br />

1. Connect the DVM HI lead to A8TP9 (+HV).<br />

2. Connect the DVM LO lead to chassis.<br />

3. Read the VWrite+ voltage shown on the sticker located on the back of A10 (the back of the<br />

display).<br />

4. Adjust A8R33 until the DVM reads within ±1 V of VWrite+. The display should be of even<br />

brightness across the screen without any smearing.<br />

5. Disconnect the DVM leads.<br />

5-12 <strong>Manual</strong> No. 20790, Rev C, November 1998<br />