Operation<br />

knows to alter its frequency correction accordingly. The PPM then uses the proper frequency correction<br />

accordingly. The PPM then uses the proper frequency correction for the detector(s) at each discrete<br />

frequency point without necessitating any operator input or any calculations (see Section 2.8.2 more<br />

information on the interfacing of external frequency signals.)<br />

To make swept frequency measurements under pulsed conditions, equipment should be set up similar to<br />

Figure 2-17. The measurement routine is initiated by pressing [MENU] (10) [F1].<br />

During the time that the fast analog output mode is active, the display window will indicate:<br />

FAST ANALOG OUTPUT<br />

MEASUREMENT MODE<br />

Press [MENU] to exit<br />

<strong>Manual</strong> No. 20790, Rev C, November 1998 2-43<br />

Superceded by Revision D, March 2009