13.11.2014 Views

Manual - 8500A Series Peak Power Meter - Giga-tronics

Manual - 8500A Series Peak Power Meter - Giga-tronics

Manual - 8500A Series Peak Power Meter - Giga-tronics

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

<strong>Series</strong> <strong>8500A</strong> <strong>Peak</strong> <strong>Power</strong> <strong>Meter</strong>s<br />

5.2.6 Delay Accuracy Test<br />

Be sure that a Pulse Generator with at least 10 ppm frequency accuracy is used for this test.<br />

Pulse Generator<br />

Signal Generator<br />

<strong>Series</strong> <strong>8500A</strong><br />

Modulation<br />

Trigger<br />

Input<br />

RF Out<br />

<strong>Peak</strong> <strong>Power</strong><br />

Sensor<br />

Figure 5-2. Delay Accuracy Test Setup<br />

1. Set the peak average value to 20 by pressing [MENU] (2) [F2] [20] [UNITS]. Set the internal<br />

delay for 75 µs by pressing [MENU] (9) [F2] [75] [µs].<br />

2. Set the Pulse Generator to generate a 10 µs wide pulse with a repetition rate of exactly 10 kHz.<br />

Adjust the amplitude to 5 V. Connect the Pulse Generator to the RF source (Signal Generator) as<br />

shown in Figure 5-2. Set the Source to generate a 50 µs pulse when triggered by the Pulse<br />

Generator. Adjust the amplitude of the Source to 0 dBm or greater. The frequency of the Source<br />

is not important. Any Source operating within the range of the PPM will be satisfactory if it can<br />

be modulated with an external pulse signal and has a rise time less than about 30 ns.<br />

3. Set the PPM to read in the Graph Mode by pressing [GRAPH]. Press [GRAPH] again to cause<br />

the autoscale function to be performed after the PPM has been connected to the Pulse Generator.<br />

A graph of the pulse profile should now be displayed.<br />

4. Select a delay window of 0.2 µs and set the cursor delay to place the cursor at the 50% point on<br />

the rising edge of the pulse. Note the Cursor Delay Value.<br />

5. Press [MENU] (4) [F1]. Set the Reference Delay to be the value read at the 50% point of the<br />

first pulse (causing the cursor delay to be zero).<br />

6. Set the start delay to 99.90 µs. Set the cursor delay to place the cursor on the 50% point of the<br />

rising edge of the pulse. The cursor delay should read 100.00 µs ±0.01 µs.<br />

5-6 <strong>Manual</strong> No. 20790, Rev C, November 1998<br />

Superceded by Revision D, March 2009

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!