RF Detectors<br />

Absolute Maximum<br />

(<strong>Peak</strong> or Avg):<br />

+23 dBm (200 mW)<br />

(Damage Limit)<br />

Return Loss (SWR):<br />

Type N, APC-7<br />

Type K<br />

Below 2 GHz 25 dB (1.12) 25 dB (1.12)<br />

2 to 12.4 GHz 20 dB (1.22) 20 dB (1.22)<br />

12.4 to 18 GHz 16 dB (1.37) 16 dB (1.37)<br />

18 to 26.5 GHz 14 dB (1.50)<br />

26.5 to 40 GHz 10 dB (1.92)<br />

Calibration Factor Uncertainty:<br />

Frequency Sum of Uncertainties (%) 1 Probable Uncertainties (%) 2<br />

Below 10 GHz 2.6% 1.2%<br />

10 to 18 GHz 6.4% 3.7%<br />

18 to 26.5 GHz 10% 6.5%<br />

26.5 to 40 GHz 20% 10%<br />

NOTES:<br />

1. Includes uncertainty of reference standard and transfer uncertainty. Directly traceable to NIST (formerly NBS).<br />

2. Square root of the sum of the individual uncertainties squared (RSS).<br />

<strong>Manual</strong> No. 20790, Rev C, November 1998 D-3<br />

Superceded by Revision D, March 2009